4.7 Article

Characterization and Correction of Beam-hardening Artifacts during Dynamic Volume CT Assessment of Myocardial Perfusion

Journal

RADIOLOGY
Volume 256, Issue 1, Pages 111-118

Publisher

RADIOLOGICAL SOC NORTH AMERICA
DOI: 10.1148/radiol.10091399

Keywords

-

Ask authors/readers for more resources

Purpose: To fully characterize beam-hardening effects caused by iodinated contrast medium in the left ventricular (LV) cavity and aorta in the assessment of myocardial perfusion at computed tomography (CT) and to validate a beam-hardening artifact correction algorithm that considers fluid-filled vessels and chambers important sources of beam hardening. Materials and Methods: The Johns Hopkins University animal care and use committee approved all procedures. An anatomically correct LV and myocardial phantom to characterize beam-hardening artifacts was designed. Following validation in the phantom, the beam-hardening correction (BHC) algorithm was applied to 256-detector row dynamic volume CT images in a canine ischemia model (n = 5) during adenosine stress, and the effect of beam hardening was determined by comparing regional dynamic volume CT perfusion metrics (myocardial upslope normalized by maximum LV blood pool attenuation) with microsphere-derived myocardial blood flow (MBF). A paired Student t test was used to compare continuous variables from the same subject but under different conditions, while linear regression analysis was performed to estimate the slope and statistical significance of the relationship between CT-derived perfusion metrics and microsphere-derived MBF. Results: Beam-hardening artifacts were successfully reproduced in phantom studies and were eliminated with the BHC algorithm. The correlation coefficient of CT-derived perfusion metrics and microsphere-derived MBF improved from 0.60 to 0.74 (P > .05) following correction in the animal model. Conclusion: Beam-hardening artifacts confound dynamic volume CT assessment of myocardial perfusion. Application of the BHC algorithm is helpful for improving accuracy of myocardial perfusion at dynamic volume CT. (C) RSNA, 2010

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available