Validation of Dual Energy X-Ray Absorptiometry Measures of Abdominal Fat by Comparison with Magnetic Resonance Imaging in an Indian Population

Journal

PLOS ONE
Volume 7, Issue 12, Pages -

Publisher

PUBLIC LIBRARY SCIENCE
DOI: 10.1371/journal.pone.0051042

Keywords

-

Funding

  1. Wellcome Trust Grant [WT083707MA]
  2. Wellcome Trust PhD Studentship [WT083431MA]
  3. UK Medical Research Council [G0600705]
  4. MRC
  5. National Institute of Nutrition
  6. MRC [MC_U120061305, G0600705] Funding Source: UKRI
  7. Medical Research Council [MC_U120061305, G0600705] Funding Source: researchfish

Ask authors/readers for more resources

Objective: Abdominal adiposity is an important risk factor for diabetes and cardiovascular disease in Indians. Dual energy Xray absorptiometry (DXA) can be used to determine abdominal fat depots, being more accessible and less costly than gold standard measures such as magnetic resonance imaging (MRI). DXA has not been fully validated for use in South Asians. Here, we determined the accuracy of DXA for measurement of abdominal fat in an Indian population by comparison with MRI. Design: 146 males and females (age range 18-74, BMI range 15-46 kg/m(2)) from Hyderabad, India underwent whole body DXA scans on a Hologic Discovery A scanner, from which fat mass in two abdominal regions was calculated, from the L1 to L4 vertebrae (L1L4) and from the L2 to L4 vertebrae (L2L4). Abdominal MRI scans (axial T1-weighted spin echo images) were taken, from which adipose tissue volumes were calculated for the same regions. Results: Intra-class correlation coefficients between DXA and MRI measures of abdominal fat were high (0.98 for both regions). Although at the level of the individual, differences between DXA and MRI could be large (95% of DXA measures were between 0.8 and 1.4 times MRI measures), at the sample level, DXA only slightly overestimated MRI measures of abdominal fat mass (mean difference in L1L4 region: 2% (95% CI: 0%, 5%), mean difference in L2L4 region: 4% (95% CI: 1%, 7%)). There was evidence of a proportional bias in the association between DXA and MRI (correlation between difference and mean 20.3), with overestimation by DXA greater in individuals with less abdominal fat (mean bias in leaner half of sample was 6% for L1L4 (95% CI: 2, 11%) and 7% for L2L4 (95% CI: 3,12%). Conclusions: DXA measures of abdominal fat are suitable for use in Indian populations and provide a good indication of abdominal adiposity at the population level. Citation: Taylor AE, Kuper H, Varma RD, Wells JC, Bell JD, et al. (2012) Validation of Dual Energy X-Ray Absorptiometry Measures of Abdominal Fat by Comparison with Magnetic Resonance Imaging in an Indian Population. PLoS ONE 7(12): e51042. doi:10.1371/journal.pone.0051042

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available