The effects of photon flux on energy spectra and imaging characteristics in a photon-counting x-ray detector

Journal

PHYSICS IN MEDICINE AND BIOLOGY
Volume 58, Issue 14, Pages 4865-4879

Publisher

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/58/14/4865

Keywords

-

Funding

  1. National Research Foundation of Korea (NRF)
  2. Korea government (MEST) [2012-0004846]

Ask authors/readers for more resources

The purpose of this paper was to investigate the effect of photon flux on the recorded energy spectrum and images produced with a photon-counting detector. We used a photon-counting cadmium telluride (CdTe) x-ray detector (model PID350, Oy Ajat, Finland). The CdTe array was composed of 16 384 pixels, each 0.35 x 0.35 x 0.75mm(3) in dimension. The photon flux is controlled by an additional aluminum filter (1, 10, 20, 30 and 40 mm). Images were acquired at three different tube voltages (50, 70 and 90 kVp) with various thicknesses of photon flux control (PFC) filters. The data acquisition time was changed to acquire an approximately equal number of counts within the selected energy window between different thicknesses of PFC filters at the same tube voltage. A phantom was manufactured to evaluate the photon flux effect on the image. The phantom was made from polymethyl methacrylate and four concentrations of iodine. The photon flux effect on the image was evaluated by the signal-difference-to-noise ratio (SDNR) between iodine and the background material. The changes of photon flux affected the recorded energy spectra and image. The thickness of the PFC filter that showed the maximum SDNR differed according to the tube voltage. The 10 mm PFC filter showed the highest SDNR at 50 and 70 kVp, while the 30 mm PFC filter exhibited the highest SDNR at 90 kVp. The SDNR was improved up to, on average, 30-fold in optimal photon flux conditions which acquired a spectrum including the lowest electronic noise with no pulse pile-up effect. The results of this study showed that the photon flux affected not only the acquired energy spectrum but also the image. Based on these results, the spectral distortion correction should be considered in connection with the image that is the ultimate purpose of medical imaging.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available