Geometrical co-calibration of a tomographic optical system with CT for intrinsically co-registered imaging

Journal

PHYSICS IN MEDICINE AND BIOLOGY
Volume 55, Issue 6, Pages 1591-1606

Publisher

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/55/6/004

Keywords

-

Ask authors/readers for more resources

A mathematical approach for geometric co-calibration of a dual-modal small-animal imaging system is presented. The system comprises an optical imaging setup for in vivo bioluminescence and fluorescence detection, as well as an x-ray CT, both mounted on a common rotatable gantry enabling fully simultaneous imaging at axially overlapping fields-of-view. Geometric co-calibration is performed once by imaging a single cylindrical light-emitting source with both modalities over 360 degrees at two axial positions, respectively. Given the three-dimensional coordinates of the source positions in the reconstructed CT volume data along with their two-dimensional locations projected at the optical detector plane, the following intrinsic system parameters are calculated: (i) the intrinsic geometric parameters of the optical detection system-five parameters for each view and (ii) the relative positional relationship between the optical and CT systems-two parameters for each view. After co-calibration is performed, experimental studies using phantoms demonstrate the high degree of intrinsic positional accuracy between the optical and CT measurements. The most important advantage of this approach is that dual-modal data fusion is accomplished without any post-registration strategies.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available