4.6 Article

Reflection-mode optical-resolution photoacoustic microscopy based on a reflective objective

Journal

OPTICS EXPRESS
Volume 21, Issue 20, Pages 24210-24218

Publisher

OPTICAL SOC AMER
DOI: 10.1364/OE.21.024210

Keywords

-

Categories

Funding

  1. National Major Scientific Research Program of China [2011CB910401]
  2. Science Fund for Creative Research Group of China [61121004]
  3. National Natural Science Foundation of China [81201067]

Ask authors/readers for more resources

We developed a new reflection-mode optical-resolution photoacoustic microscopy (OR-PAM) based on the cooperation of a reflective objective and an ultrasonic transducer. The reflective objective is used to produce nearly diffraction-limited optical focusing, and the excited ultrasound waves are then directly detected by an ultrasonic transducer that was placed in the central cone of the objective. This new design avoids the coupling between optical focusing and ultrasound transmission in the reflection mode. Moreover, the proposed system is able to provide lateral resolution of 1.2 mu m at 580 nm, penetration depth of 0.9 mm in biological tissues, and a work distance of 6.0 mm. We present in vivo imaging of the microvasculature in mouse ears and in vitro imaging of red blood cells (RBCs), which demonstrate the capability of the system to study microcirculation. (C) 2013 Optical Society of America