4.4 Article

Performance of cerium-doped Gd3Al2Ga3O12 (GAGG:Ce) scintillator in gamma-ray spectrometry

Publisher

ELSEVIER
DOI: 10.1016/j.nima.2013.01.064

Keywords

Scintillation detectors; Ce-doped crystals; Nonproportionality; Fast timing; SiPMs light readout

Ask authors/readers for more resources

Performance of cerium-doped Gd3Al2Ga3O12 (GAGG:Ce) scintillator in gamma-ray spectrometry has been investigated. The measurements of two samples of GAGG:Ce cover the tests of emission spectra (maximum of emission at about 530 nm), light output, non-proportionality, energy resolution, time resolution and decay time of light pulses. We compare the results with commonly known scintillators, such as NaI(Tl), LSO, LuAG etc. The results show that GAGG:Ce has a high light yield of about 33000 ph/MeV as measured with Hamamatsu S3590-18 Si PiN photodiode [1]. The total energy resolution for 662 keV gamma-rays from Cs-137 source is equal to about 6%, whereas intrinsic resolution is equal to 5.2%. Additionally, we made basic measurements of photoelectron yield, non-proportionality and total energy resolution of small sample (5 x 5 x 5 mm(3)) of GAGG:Ce crystal coupled to Hamamatsu MPPC array (6 x 6 mm(2)). The results show that the performance of GAGG:Ce measured with the MPPC array are similar to those measured with the PMT. (c) 2013 Elsevier B.V. All rights reserved.