4.4 Article

A practical method to determine the spatial resolution of GEM detector

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/j.nima.2012.10.078

Keywords

Gaseous electron multiplication; Spatial resolution

Funding

  1. National Natural Science Foundation of China [10975083, 11079025, 11120101004]
  2. Tsinghua University Initiative Scientific Research Program

Ask authors/readers for more resources

A new method to determine the spatial resolution of a one-dimensional gaseous electron multiplication (GEM) prototype is demonstrated via an experimental test using an Fe-55 source with an adjustable slit. In this method, the total variance of the incident position sigma(2)(tot), contributed by the GEM spatial resolution and the geometrical acceptance of the slit, is presented as a function of the event rate n under the full energy peak, with the latter being employed to calibrate the square of the slit width. Using a linear fit to the experimental data, a spatial resolution of sigma(g) = 56 +/- 15 mu m (1 sigma) is measured. Compared to the existing methods, the novel method presented here is robust considering the inaccuracy in the direct measurements of the slit widths, and hence greatly circumvents machining difficulties. (C) 2012 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available