In situ monitoring of structural changes during the adsorption on flexible porous coordination polymers by X-ray powder diffraction: Instrumentation and experimental results

Journal

MICROPOROUS AND MESOPOROUS MATERIALS
Volume 188, Issue -, Pages 190-195

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/j.micromeso.2013.12.024

Keywords

ELM-11; In-situ adsorption; Powder X-ray diffraction; Flexible MOFs

Funding

  1. Federal Ministry of Education and Research (BMBF) [05K100D3]
  2. Helmhotz Zentrum Berlin far Materialien und Energie
  3. Nippon Steel Corporation, Futtsu, Japan [ELM-11]

Ask authors/readers for more resources

A new instrumentation for in situ X-ray powder diffraction allows to follow structural changes in the crystal structure of flexible coordination polymers during adsorption and desorption of gases. The system is based on closed cycle He-cryostat that ensures the high temperature stability in the range of 5.5-450 K. The automated gas dosing system facilitates the physisorption measurements in the pressure range of 0.01-130 kPa. The system is implemented on KMC-2 and MAGS beamlines at Helmholtz Zentrum Berlin fur Materialien und Energie. Using new instrumentation, the adsorption of CO2 and n-butane at 273 K on a flexible two-dimensional coordination polymer ELM-11 (ELM - Elastic Layer Material) was investigated. (C) 2014 Elsevier Inc. All rights reserved.