4.6 Article

Report of AAPM TG 135: Quality assurance for robotic radiosurgery

Journal

MEDICAL PHYSICS
Volume 38, Issue 6, Pages 2914-2936

Publisher

WILEY
DOI: 10.1118/1.3579139

Keywords

quality assurance; stereotactic radiosurgery; radiation therapy; robotic radiosurgery

Ask authors/readers for more resources

The task group (TG) for quality assurance for robotic radiosurgery was formed by the American Association of Physicists in Medicine's Science Council under the direction of the Radiation Therapy Committee and the Quality Assurance (QA) Subcommittee. The task group (TG-135) had three main charges: (1) To make recommendations on a code of practice for Robotic Radiosurgery QA; (2) To make recommendations on quality assurance and dosimetric verification techniques, especially in regard to real-time respiratory motion tracking software; (3) To make recommendations on issues which require further research and development. This report provides a general functional overview of the only clinically implemented robotic radiosurgery device, the CyberKnife (R). This report includes sections on device components and their individual component QA recommendations, followed by a section on the QA requirements for integrated systems. Examples of checklists for daily, monthly, annual, and upgrade QA are given as guidance for medical physicists. Areas in which QA procedures are still under development are discussed. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3579139]

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available