Dosimetric verification of the anisotropic analytical algorithm in lung equivalent heterogeneities with and without bone equivalent heterogeneities

Journal

MEDICAL PHYSICS
Volume 37, Issue 8, Pages 4456-4463

Publisher

AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
DOI: 10.1118/1.3464748

Keywords

dosimetric verification; AAA dose calculation algorithm; Monte Carlo; MCNP; convolution-superposition algorithm; heterogeneity corrections

Ask authors/readers for more resources

Purpose: In this study, the authors evaluated the accuracy of dose calculations performed by the convolution/superposition based anisotropic analytical algorithm (AAA) in lung equivalent heterogeneities with and without bone equivalent heterogeneities. Methods: Calculations of PDDs using the AAA and Monte Carlo simulations (MCNP4C) were compared to ionization chamber measurements with a heterogeneous phantom consisting of lung equivalent and bone equivalent materials. Both 6 and 10 MV photon beams of 4 x 4 and 10 x 10 cm(2) field sizes were used for the simulations. Furthermore, changes of energy spectrum with depth for the heterogeneous phantom using MCNP were calculated. Results: The ionization chamber measurements and MCNP calculations in a lung equivalent phantom were in good agreement, having an average deviation of only 0.64 +/- 0.45%. For both 6 and 10 MV beams, the average deviation was less than 2% for the 4 x 4 and 10 x 10 cm(2) fields in the water-lung equivalent phantom and the 4 x 4 cm(2) field in the water-lung-bone equivalent phantom. Maximum deviations for the 10 x 10 cm(2) field in the lung equivalent phantom before and after the bone slab were 5.0% and 4.1%, respectively. The Monte Carlo simulation demonstrated an increase of the low-energy photon component in these regions, more for the 10 x 10 cm(2) field compared to the 4 x 4 cm(2) field. Conclusions: The low-energy photon by Monte Carlo simulation component increases sharply in larger fields when there is a significant presence of bone equivalent heterogeneities. This leads to great changes in the build-up and build-down at the interfaces of different density materials. The AAA calculation modeling of the effect is not deemed to be sufficiently accurate. (C) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3464748]

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available