4.6 Article

An easy-to-use phantom and protocol for weekly PET quality assessment: A multicenter study

Journal

MEDICAL PHYSICS
Volume 35, Issue 9, Pages 3922-3934

Publisher

WILEY
DOI: 10.1118/1.2964091

Keywords

PET; quality assessment; phantom; image processing

Funding

  1. Institut National du Cancer (INCa, CancEropole Nord-Ouest)

Ask authors/readers for more resources

The authors have developed a simple phantom and dedicated software for the quality assessment of positron emission tomography (PET) scanners. The phantom is a parallelepiped box filled with a relatively low activity (18)FDG solution and in which simple test objects are placed. Various image quality parameters are checked, including signal-to-noise ratio, image uniformity, slice thickness, slice sensitivity profile, spatial resolution, and dose calibration accuracy. Automatic image analysis consists in detecting surfaces and objects, defining regions of interest, acquiring reference point coordinates, and establishing gray-scale profiles. The total time needed for quality assessment (preparation and image acquisition) is less than 15 min with 37 MBq (1 mCi) (18)FDG. The system's ease of use encourages frequent image quality assessment-for example, the comparison of PET scanners in interdepartment studies and the monitoring and evaluation of possible drifts over time. By way of an example, the authors present weekly quality assessment results obtained over up to 7 months at four PET facilities. (C) 2008 American Association of Physicists in Medicine.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available