4.1 Article

DOSIMETRIC COMPARISON OF SPLIT FIELD AND FIXED JAW TECHNIQUES FOR LARGE IMRT TARGET VOLUMES IN THE HEAD AND NECK

Journal

MEDICAL DOSIMETRY
Volume 36, Issue 1, Pages 6-9

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.meddos.2009.10.002

Keywords

IMRT; Fixed-jaw treatment; Large field; Dosimetry

Ask authors/readers for more resources

Some treatment planning systems (TPSs), when used for large-field (>14 cm) intensity-modulated radiation therapy (IMRT), create split fields that produce excessive multiple-leaf collimator segments, match-line dose inhomogeneity, and higher treatment times than nonsplit fields. A new method using a fixed-jaw technique (FJT) forces the jaw to stay at a fixed position during optimization and is proposed to reduce problems associated with split fields. Dosimetric comparisons between split-field technique (SFT) and FJT used for IMRT treatment is presented. Five patients with head and neck malignancies and regional target volumes were studied and compared with both techniques. Treatment planning was performed on an Eclipse TPS using beam data generated for Varian 2100C linear accelerator. A standard beam arrangement consisting of nine coplanar fields, equally spaced, was used in both techniques. Institutional dose-volume constraints used in head and neck cancer were kept the same for both techniques. The dosimetric coverage for the target volumes between SFT and FJT for head and neck IMRT plan is identical within +/- 1% up to 90% dose. Similarly, the organs at risk (OARs) have dose-volume coverage nearly identical for all patients. When the total monitor unit (MU) and segments were analyzed, SFT produces statistically significant higher segments (17.3 +/- 6.3%) and higher MU (13.7 +/- 4.4%) than the FJT. There is no match line in FJT and hence dose uniformity in the target volume is superior to the SET. Dosimetrically, SFT and FJT are similar for dose-volume coverage; however, the FJT method provides better logistics, lower MU, shorter treatment time, and better dose uniformity. The number of segments and MU also has been correlated with the whole body radiation dose with long-term complications. Thus, FJT should be the preferred option over SFT for large target volumes. (C) 2011 American Association of Medical Dosimetrists.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.1
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available