4.7 Article Proceedings Paper

Experimental complex for investigations of high temperature capillarity phenomena

Publisher

ELSEVIER SCIENCE SA
DOI: 10.1016/j.msea.2007.11.094

Keywords

High-temperature tests; Wettability; Reactivity; Equipment design

Ask authors/readers for more resources

The paper presents the description of an experimental complex that has been designed for investigations of high temperature capillarity phenomena by various testing methods (classical sessile drop, pendant drop, dispensed drop, sandwiched drop, transferred drop, drop sucking, drop pushing, drop smearing or rubbing) at a temperature of up to 2100 degrees C under vacuum of up to 10(-7) hPa or in protective atmosphere (static or flowing gas with controlled rate at required level of pressure). Several examples of high temperature wettability tests are discussed in order to demonstrate the wide testing possibilities of the new experimental complex. (C) 2008 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available