4.3 Article

Gold nanoparticles modified carbon paste electrode for differential pulse voltammetric determination of eugenol

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/j.msec.2014.06.035

Keywords

Gold nanoparticles; Modified carbon paste electrode; Eugenol determination; Differential pulse voltammetry

Ask authors/readers for more resources

In the present study, a carbon paste electrode chemically modified with gold nanoparticles was used as a sensitive electrochemical sensor for determination of eugenol. The differential pulse voltammetric method was employed to study the behavior of eugenol on this modified electrode. The effect of variables such as percent of gold nanoparticles, pH of solution, accumulation potential and time on voltammogram peak current were optimized. The proposed electrode showed good oxidation response for eugenol in 0.1 mol L-1 phosphate buffer solution (pH 8) and the peak potential was about +285 mV (vs. Ag/AgCl). The peak current increased linearly with the eugenol concentration in the range of 5-250 mu mol L-1. The detection limit was found to be 2.0 mu mol L-1 and the relative standard deviation was 1.2% (n = 7). The effect of interferences on the eugenol peak current was studied. The method has been applied to the determination of eugenol in different real samples, spiked recoveries were in the range of 96%-99%. (C) 2014 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available