4.5 Article

Prospective Head-Movement Correction for High-Resolution MRI Using an In-Bore Optical Tracking System

Journal

MAGNETIC RESONANCE IN MEDICINE
Volume 62, Issue 4, Pages 924-934

Publisher

WILEY
DOI: 10.1002/mrm.22076

Keywords

prospective motion correction; optical tracking; MR-compatible camera; real-time MRI; high-resolution MRI

Funding

  1. Intramural NIH HHS [Z01 NS002990-08] Funding Source: Medline

Ask authors/readers for more resources

In MRI of the human brain, subject motion is a major cause of magnetic resonance image quality degradation. To compensate for the effects of head motion during data acquisition, an in-bore optical motion tracking system is proposed. The system comprises two MR-compatible infrared cameras that are fixed on a holder right above and in front of the head coil. The resulting close proximity of the cameras to the object allows precise tracking of its movement. During image acquisition, the MRI scanner uses this tracking information to prospectively compensate for head motion by adjusting the gradient field direction and radio frequency (RF) phases and frequencies. Experiments performed on subjects demonstrate robust system performance with translation and rotation accuracies of 0.1 mm and 0.15 degrees, respectively. Magn Reson Med 62:924-934, 2009. (C) 2009 Wiley-Liss, Inc.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available