Rapid Single-Scan T2*-Mapping Using Exponential Excitation Pulses and Image-Based Correction for Linear Background Gradients

Journal

MAGNETIC RESONANCE IN MEDICINE
Volume 62, Issue 1, Pages 263-268

Publisher

JOHN WILEY & SONS INC
DOI: 10.1002/mrm.21971

Keywords

effective transverse relaxation time; T(2)*; susceptibility artifact; field inhomogeneity; gradient echo

Funding

  1. Bundesministerium fur Bildung und Forschung (Brain Imaging Center Frankfurt) [DLR 01GO0203]
  2. Deutsche Forschungsgemeinschaft [ZA 233/1-1]
  3. Interdisciplinary Center of Neuroscience (ICN), Frankfurt am Main

Ask authors/readers for more resources

A method for fast quantitative T(2)* mapping based on multiple gradient-echo (multi-GE) imaging with correction for static magnetic field inhomogeneities is described, using an exponential excitation pulse. Field gradient maps are obtained from the phase information and modulus data are subsequently corrected, allowing for simple monoexponential T(2)* fitting. Echoes with long echo times suffering from major signal losses due to field inhomogeneities are excluded from the analysis. The acquisition time for a matrix size of 256 x 256, 1 mm in-plane resolution, and 2 mm slice thickness amounts to 15 s per slice. An additional correction for in-plane field gradients further improves accuracy. Phantom experiments show that the method provides accurate T(2)* values for field gradients up to 200 mu T/m; for gradients up to 300 mu T/m errors do not exceed 15%. In vivo T(2)* values acquired on healthy volunteers at 3T are in excellent agreement with results from the literature. Magn Reson Med 62:263-268, 2009. (C) 2009 Wiley-Liss, Inc.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available