Journal
MAGNETIC RESONANCE IN MEDICINE
Volume 61, Issue 3, Pages 626-633Publisher
WILEY
DOI: 10.1002/mrm.21499
Keywords
field inhomogeneity; MR susceptometry; oxygen saturation; phase image; retrospective correction
Funding
- National Institutes of Health [T32 EB000814, R01 HL 75649]
Ask authors/readers for more resources
MR susceptometry-based blood oximetry relies on phase mapping to measure the difference in magnetic susceptibility between intravascular blood and surrounding tissue. The main source of error in MR susceptometry is the static field inhomogeneity caused by an interface between air and tissue or between adjacent tissue types. High-pass filtering has previously been used in conjunction with shimming to reduce the effect of low spatial-frequency modulations of the phase caused by large-scale induced magnetic fields. We demonstrate that high-pass filtering is not optimum for MR susceptometry because the results are sensitive to filter size. We propose an alternative method that acquires data without scanner-implemented default shimming, and fits, after appropriate weighting and masking, the static field inhomogeneity to a second-order polynomial. Compared to shimming the retrospective correction technique improved agreement between hemoglobin saturations measured in different segments of a vessel (femoral versus popliteal artery and vein) from three standard errors to less than one. Magn Reson Med 61:626-633, 2008. (C) 2008 Wiley-Liss, Inc.
Authors
I am an author on this paper
Click your name to claim this paper and add it to your profile.
Reviews
Recommended
No Data Available