4.6 Article

Biocompatible Graphene Oxide-Based Glucose Biosensors

Journal

LANGMUIR
Volume 26, Issue 9, Pages 6158-6160

Publisher

AMER CHEMICAL SOC
DOI: 10.1021/la100886x

Keywords

-

Funding

  1. Zhejiang Department of Education [Y200906587]
  2. Ministry of Science and Technology of China [2009DFB30380]
  3. Zhejiang Department of Science and Technology [2009C13019]

Ask authors/readers for more resources

This letter demonstrates that a novel, highly efficient enzyme electrode can be directly obtained using covalent attachment between carboxyl acid groups of graphene oxide sheets and amines of glucose oxidase. The resulting biosensor exhibits a broad linear range up to 28 mM.mm(-2) glucose with a sensitivity of 8.045 mA.cm(-2), M-1. The glucose oxidase-immobilized graphene oxide electrode also shows a reproducibility and a good storage stability, suggesting potentials for a wide range of practical applications. The biocompatibility of as-synthesized graphene oxide nanosheets with human cells, especially retinal pigment epithelium (RPE) cells, was in for the first time in the present work. Microporous graphene oxide exhibits good biocompatibility and has potential advantages with respect to cell attachment and proliferation, leading to opportunities for using graphene-based biosensors for the clinical diagnosis.