Biosensor system-on-a-chip including CMOS-based signal processing circuits and 64 carbon nanotube-based sensors for the detection of a neurotransmitter

Journal

LAB ON A CHIP
Volume 10, Issue 7, Pages 894-898

Publisher

ROYAL SOC CHEMISTRY
DOI: 10.1039/b916975j

Keywords

-

Funding

  1. NANO-Systems Institute National Core Research Center (NSI-NCRC)
  2. KOSEF [20090079103]
  3. MKE
  4. National Research Foundation of Korea [2009-0081999]

Ask authors/readers for more resources

We developed a carbon nanotube (CNT)-based biosensor system-on-a-chip (SoC) for the detection of a neurotransmitter. Here, 64 CNT-based sensors were integrated with silicon-based signal processing circuits in a single chip, which was made possible by combining several technological breakthroughs such as efficient signal processing, uniform CNT networks, and biocompatible functionalization of CNT-based sensors. The chip was utilized to detect glutamate, a neurotransmitter, where ammonia, a byproduct of the enzymatic reaction of glutamate and glutamate oxidase on CNT-based sensors, modulated the conductance signals to the CNT-based sensors. This is a major technological advancement in the integration of CNT-based sensors with microelectronics, and this chip can be readily integrated with larger scale lab-on-a-chip (LoC) systems for various applications such as LoC systems for neural networks.