3.9 Article

Evaluation of Two Commercial HLA-B27 Real-Time PCR Kits

Journal

KOREAN JOURNAL OF LABORATORY MEDICINE
Volume 29, Issue 6, Pages 589-593

Publisher

KOREAN SOC LABORATORY MEDICINE
DOI: 10.3343/kjlm.2009.29.6.589

Keywords

Real Time PCR; HLA-B27; PCR-SSP

Ask authors/readers for more resources

Background: The standard PCR with sequence-specific primers (SSP) is a widely used method of HLA-B27 typing in clinical practice. The aim of our study was to evaluate 2 Korean HLA-B27 kits with different real-time PCR chemistries. Methods: To validate the accuracy of real-time PCR kits, we selected 28 HLA-B27-positive samples and 33 HLA-B27-negative samples with a wide range of different HLA-B specificities typed by standard PCR-SSP. The 2 real-time PCR kits used were the AccuPower(R) HLA-B27 real-time PCR kit (Bioneer, Korea) with TaqMan probes and the Real-Q(TM) HLA-B*27 detection kit (BioSewoom, Korea) with SYBR Green I dye for melting curve analysis. Results: All 61 samples typed by PCR-SSP demonstrated a perfect concordance with the 2 real-time PCR assays. It was possible to clearly discriminate between HLA-B27-positive and -negative samples in both real-time assays. Conclusions: In summary, both real-time PCR assays for HLA-B27 were fast, reliable, well-adapted for routine laboratory testing, and attractive alternatives to the conventional PCR-SSP method. (Korean J Lab Med 2009;29:589-93)

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

3.9
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available