4.4 Article

Comparison of Xpert Flu rapid nucleic acid testing with rapid antigen testing for the diagnosis of influenza A and B

Journal

JOURNAL OF VIROLOGICAL METHODS
Volume 186, Issue 1-2, Pages 137-140

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jviromet.2012.07.023

Keywords

Influenza A; Influenza B; Rapid antigen test; Rapid influenza diagnostic test; Direct fluorescent antibody; GeneXpert; Xpert Flu

Funding

  1. Cepheid

Ask authors/readers for more resources

Influenza infections are associated with thousands of hospital admissions and deaths each year. Rapid detection of influenza is important for prompt initiation of antiviral therapy and appropriate patient triage. In this study the Cepheid Xpert Flu assay was compared with two rapid antigen tests, BinaxNOW Influenza A & B and BD Directigen EZ Flu A + B, as well as direct fluorescent antibody testing for the rapid detection of influenza A and B. Using real-time, hydrolysis probe-based, reverse transcriptase PCR as the reference method, influenza A sensitivity was 97.3% for Xpert Flu, 95.9% for direct fluorescent antibody testing, 62.2% for BinaxNOW, and 71.6% for BD Directigen. Influenza B sensitivity was 100% for Xpert Flu and direct fluorescent antibody testing, 54.5% for BinaxNOW, and 48.5% for BD Directigen. Specificity for influenza A was 100% for Xpert Flu, BinaxNOW, and BD Directigen, and 99.2% for direct fluorescent antibody testing. All methods demonstrated 100% specificity for influenza B. These findings support the use of the Xpert Flu assay in settings requiring urgent diagnosis of influenza A and B. (C) 2012 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available