4.6 Article

Bovine pericardial versus porcine stented replacement aortic valves: Early results of a randomized comparison of the Perimount and the Mosaic valves

Journal

JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY
Volume 136, Issue 5, Pages 1142-1148

Publisher

MOSBY-ELSEVIER
DOI: 10.1016/j.jtcvs.2007.12.086

Keywords

-

Funding

  1. Medtronic

Ask authors/readers for more resources

Objective: A stented bovine pericardial valve might be less obstructive than a stented porcine valve. This study compared early hemodynamic function in a prospective series of 99 patients randomized to receive either a Mosaic or Perimount replacement aortic valve. Methods: Echocardiography was performed early after surgery and at 1 year after surgery. Patients also filled in psychologic questionnaires and underwent a 6-minute walk. Results: The groups were matched demographically. The Perimount valve was significantly less obstructive in terms of mean pressure difference (11 6 5 vs 17 6 7 mm Hg; P<.0001), with a trend in favor of a larger effective orifice area (1.47 +/- 6 0.45 vs 1.28 +/- 6 0.46 cm(2); P = 5.05) postoperatively. There were no differences in left ventricular mass regression, aortic regurgitation, 6-minute walk, psychologic questionnaires, or mortality and clinical events. Conclusion: The stented bovine pericardial valve was less obstructive than the stented porcine valve. Both valves were associated with similar and significant improvements in quality of life, exercise ability, and regression of left ventricular mass.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available