4.2 Article

The MYTHEN detector for X-ray powder diffraction experiments at the Swiss Light Source

Journal

JOURNAL OF SYNCHROTRON RADIATION
Volume 17, Issue -, Pages 653-668

Publisher

INT UNION CRYSTALLOGRAPHY
DOI: 10.1107/S0909049510026051

Keywords

detectors; powder diffraction

Ask authors/readers for more resources

The MYTFIEN single-photon-counting silicon microstrip detector has been developed at the Swiss Light Source for time-resolved powder diffraction experiments An upgraded version of the detector has been installed at the SLS powder diffraction station allowing the acquisition of diffraction patterns over 120 degrees in 20 in fractions of seconds. Thanks to the outstanding performance of the detector and to the calibration procedures developed, the quality of the data obtained IS now comparable with that of traditional high-resolution point detectors in terms of FWHM resolution and peak profile shape. with the additional advantage of fast and simultaneous acquisition of the full diffraction pattern MYTH EN is therefore optimal for time-resolved or dose-critical measurements The characteristics of the MYTH EN detector together with the calibration procedures Implemented for the optimization of the data are described in detail The refinements of two known standard powders are discussed together with a remarkable application of MYTHEN to organic compounds in relation to the problem of radiation damage.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available