4.5 Article

Fast and simple method for the determination of urinary 1-hydroxypyrene

Journal

JOURNAL OF SEPARATION SCIENCE
Volume 32, Issue 20, Pages 3403-3410

Publisher

WILEY-V C H VERLAG GMBH
DOI: 10.1002/jssc.200900382

Keywords

Liquid chomatography; Polycyclic aromatic hydrocarbons; Urinary pyrene metabolites

Ask authors/readers for more resources

A fast method for the measurement of metabolites of pyrene in urine was improved by HPLC with fluorescence detector using heart-cut technique. This method can quantify the total amount of pyrene metabolites corresponding to glucuronic acid and sulfate conjugates as well as free 1-OH-Py. The hydrolyzed biological fluid was directly injected into the chromatographic system, via a column-switching system. Pre-treatment and analysis were performed within 0.5 and 9 min, respectively. Enzymatic hydrolysis has been optimized to not exceed 2 h. The best response function, in the 0.2-10 ng/mL range, is the linear regression, bringing simplicity, good accuracy, and low LOQ The intra- and interday precision values were inferior to 1 and 2%, respectively. The proposed method provided a simple, convenient, and practical procedure to determine the level of the main urinary pyrene metabolites in biological samples.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available