4.6 Article

LC-determination of five paraben preservatives in saliva and toothpaste samples using UV detection and a short monolithic column

Journal

JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS
Volume 53, Issue 3, Pages 785-789

Publisher

ELSEVIER
DOI: 10.1016/j.jpba.2010.05.018

Keywords

High-performance liquid chromatography; Monolithic column; Parabens; Saliva; Toothpaste; Solid phase extraction

Ask authors/readers for more resources

The present study reports the development and application of an HPLC-UV method for the simultaneous separation and determination of five paraben preservatives (methyl-, ethyl-, propyl-, n-butyl- and iso-butyl-paraben) in real samples All analytes were separated efficiently in less than 20 min using a simple H2O center dot ACN linear gradient and a short monolithic column (50 mm x 4.6 min i.d) at a flow rate of 3 0 mL min(-1). Phenoxyethanol was used as chromatographic internal standard The method was validated for linearity, limits of detection and quantification, accuracy and precision Human saliva and toothpaste samples were analyzed after SPE pretreatment on Licrolut RP-18 cartridges The detection limits varied between 0.1 and 0 3 mg L-1 in all cases and the percent recoveries between 86 and 113% (C) 2010 Elsevier B V All rights reserved

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available