4.6 Article Proceedings Paper

A new approach to the application of solid phase extraction disks with LC-MS/MS for the analysis of drugs on a 96-well plate format

Journal

JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS
Volume 50, Issue 4, Pages 556-562

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jpba.2008.07.014

Keywords

96-Well disk solid phase extraction; Liquid chromatography; Automation; Benzodiazepines; Drug analysis

Ask authors/readers for more resources

A new 96-well disk solid phase extraction sample preparation technique which does not involve vacuum pumps integrated with liquid chromatographic tandem mass spectrometric (LC-MS/MS) was developed for high throughput determination of benzodiazepines (nordiazepam, diazepam, lorazepam and oxazepam). In addition, the method completely allows the re-use of the SPE disk membranes for subsequent analyses after re-conditioning. The method utilizes a robotic autosampler for parallel extractions in a 96-well plate format. Results have been presented for independent extractions from three matrices; phosphate buffer solution, urine, and plasma. Factors affecting data reproducibility, extraction kinetics, sample throughput, and reliability of the system were investigated and optimized. A total time required per sample was 0.94 min using 96-well format. Method reproducibility was <= 9% relative standard deviation for all three matrices. Limits of detection and quantitation recorded were respectively in the range 0.02-0.15 and 0.2-2.0 ng/mL with linearity ranging from 0.2 to 500 ng/mL for all matrices. (C) 2008 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available