4.4 Article

High-resolution fMRI: Overcoming the signal-to-noise problem

Journal

JOURNAL OF NEUROSCIENCE METHODS
Volume 178, Issue 2, Pages 357-365

Publisher

ELSEVIER
DOI: 10.1016/j.jneumeth.2008.12.011

Keywords

Functional MRI; Structure adaptive smoothing; High-resolution fMRI

Funding

  1. Cervical Spine Research Society
  2. Institute for Biomedical Imaging Sciences (IBIS)

Ask authors/readers for more resources

Increasing the spatial resolution in functional Magnetic Resonance Imaging (fMRI) inherently lowers the signal-to-noise ratio (SNR). In order to still detect functionally significant activations in high-resolution images, spatial smoothing of the data is required. However, conventional non-adaptive smoothing comes with a reduced effective resolution, foiling the benefit of the higher acquisition resolution. We show how our recently proposed structural adaptive smoothing procedure for functional MRI data can improve signal detection of high-resolution fMRI experiments regardless of the lower SNR. The procedure is evaluated on human visual and sensory-motor mapping experiments. In these applications, the higher resolution could be fully utilized and high-resolution experiments were outperforming normal resolution experiments by means of both statistical significance and information content. (C) 2008 Elsevier B.V. All rights reserved