4.2 Article

Improved solid-state planar Ti/Pd/Ag/AgCl/KCl-gel microreference electrode by silicon cap sealing package

Journal

Publisher

SPIE-SOC PHOTO-OPTICAL INSTRUMENTATION ENGINEERS
DOI: 10.1117/1.3196547

Keywords

Chip-level packaging; Ti/Pd/Ag/AgCl/KCl-gel micro reference electrode; drift and offset voltage; impedance and phase shift; reproducibility

Funding

  1. National Science Council of Taiwan [NSC 92-2212-E-110-020]

Ask authors/readers for more resources

Although solid-state microreference electrodes (mu REs) have been developed for electrochemical and biomedical sensing applications for more than one decade, there are very few studies devoted to improving their performance through packaging. Using planar technology and a silicon cap sealing method, we implement a packaged Ti/Pd/Ag/AgCl/KCl-gel mu RE with a total dimension of only 9 mm (L) x 6 mm (W) x 1 mm (H), a hundredfold less than the commercial Ag/AgCl reference electrode (RE). Compared with the unpackaged mu RE, the presented chip-level packaged mu RE demonstrates many improved characteristics, including a very stable cell potential (5-mV drift voltage in 30,000 s), an approximately zero offset voltage (-7 mV), a very low impedance (1.50 k Omega) and phase shift (8.98 deg) at 1-kHz operation frequency, a very small double-layer capacitance (0.04 mu F), and a very high reproducibility (+/- 3.1-mV). (C) 2009 Society of Photo-Optical Instrumentation Engineers. [DOI: 10.1117/1.3196547]

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available