4.5 Article

Highly sensitive pyrogen detection on medical devices by the monocyte activation test

Journal

JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE
Volume 25, Issue 4, Pages 1065-1075

Publisher

SPRINGER
DOI: 10.1007/s10856-013-5136-6

Keywords

-

Ask authors/readers for more resources

Pyrogens are components of microorganisms, like bacteria, viruses or fungi, which can induce a complex inflammatory response in the human body. Pyrogen contamination on medical devices prior operation is still critical and associated with severe complications for the patients. The aim of our study was to develop a reliable test, which allows detection of pyrogen contamination on the surface of medical devices. After in vitro pyrogen contamination of different medical devices and incubation in a rotation model, the human whole blood monocyte activation test (MAT), which is based on an IL-1 beta-specific ELISA, was employed. Our results show that when combining a modified MAT protocol and a dynamic incubation system, even smallest amounts of pyrogens can be directly detected on the surface of medical devices. Therefore, screening of medical devices prior clinical application using our novel assay, has the potential to significantly reduce complications associated with pyrogen-contaminated medical devices.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available