4.7 Article

Automated liver stiffness measurements with magnetic resonance elastography

Journal

JOURNAL OF MAGNETIC RESONANCE IMAGING
Volume 38, Issue 2, Pages 371-379

Publisher

WILEY
DOI: 10.1002/jmri.23980

Keywords

MR elastography; liver; automation; segmentation; hepatic fibrosis

Funding

  1. National Institutes of Health (NIH) [EB01981]

Ask authors/readers for more resources

Purpose: To provide a fully automated algorithm for obtaining stiffness measurements from hepatic magnetic resonance elastography (MRE) images that are consistent with measurements performed by expert readers. Materials and Methods: An initial liver contour was found using an adaptive threshold and expanded using an active contour to select a homogeneous area of the liver. The confidence map generated during the stiffness calculation was used to select a region of reliable wave propagation. The average stiffness within the automatically generated region of interest (ROI) was compared to measurements by two trained readers in a set of 88 clinical test cases ranging from healthy to severely fibrotic. Results: The stiffness measurements reported by the readers differed by -6.76% +/- 22.8% (95% confidence) and had an intraclass correlation coefficient (ICC) of 0.972 (P < 0.05). The algorithm and the more experienced reader differed by 4.32% +/- 14.9 with an ICC of 0.987. Conclusion: The automated algorithm performed reliably, even though MRE acquisitions often have motion artifacts present. The correlation between the automated measurements and those from the trained readers was superior to the correlation between the readers. J. Magn. Reson. Imaging 2013;38:371-379. (c) 2013 Wiley Periodicals, Inc.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available