Improved slice selection for R2*mapping during cryoablation with eddy current compensation

Journal

JOURNAL OF MAGNETIC RESONANCE IMAGING
Volume 28, Issue 1, Pages 190-198

Publisher

WILEY
DOI: 10.1002/jmri.21396

Keywords

eddy current correction; ultra short TE imaging; half pulse excitation; cryoablation; temperature mapping

Ask authors/readers for more resources

Purpose: To improve the slice profile and image quality of R2* mapping in the iceball during cryoablation with ultra-short echo time (UTE) imaging by compensating for eddy currents induced by the selective gradient when half-pulse radiofrequency (RF) excitation is employed to achieve UTEs. Materials and Methods: A method to measure both 130 and linear eddy currents simultaneously is first presented. This is done with a least-square fitting process on calibration data collected on a phantom. Eddy currents during excitation are compensated by redesigning the RF pulse and the selective gradient accordingly, while that resultant from the readout gradient are compensated for during image reconstruction. In vivo data were obtained continuously during the cryoablation experiments to calculate the R2* values in the iceball and to correlate them with the freezing process. Results: Image quality degradation due to eddy currents is significantly reduced with the proposed approaches. R2* maps of iceball throughout the cryoablation experiments were achieved with improved quality. Conclusion: The proposed approaches are effective for compensating eddy currents during half-pulse RF excitation as well as readout. TEs as short as 100 mu sec were obtained, allowing R2* maps to be obtained from frozen tissues with improved quality.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available