4.3 Article Proceedings Paper

The Medipix3RX: a high resolution, zero dead-time pixel detector readout chip allowing spectroscopic imaging

Journal

JOURNAL OF INSTRUMENTATION
Volume 8, Issue -, Pages -

Publisher

IOP PUBLISHING LTD
DOI: 10.1088/1748-0221/8/02/C02016

Keywords

Analogue electronic circuits; Electronic detector readout concepts (solid-state); Digital electronic circuits

Ask authors/readers for more resources

The Medipix3 chips have been designed to permit spectroscopic imaging in highly segmented hybrid pixel detectors. Spectral degradation due to charge sharing in the sensor has been addressed by means of an architecture in which adjacent pixels communicate in the analog and digital domains on an event-by-event basis to reconstruct the deposited charge in a neighbourhood prior to the assignation of the hit to a single pixel. The Medipix3RX chip architecture is presented. The first results for the characterization of the chip with 300 m m thick Si sensors are given. similar to 72e(-) r.m.s. noise and similar to 40e(-) r.m.s. of threshold dispersion after chip equalization have been measured in Single Pixel Mode of operation. The homogeneity of the image in Charge Summing mode is comparable to the Single Pixel Mode image. This demonstrates both modes are suitable for X-ray imaging applications.