4.5 Article

Fatigue Life of Reciproc and Mtwo Instruments Subjected to Static and Dynamic Tests

Journal

JOURNAL OF ENDODONTICS
Volume 39, Issue 5, Pages 693-696

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.joen.2012.11.048

Keywords

Dynamic test; fatigue fracture; fatigue resistance; flexibility; reciprocating movement

Ask authors/readers for more resources

Introduction: This study evaluated the influence of flexibility and reciprocating movement on the fatigue life of endodontic instruments subjected to static and dynamic tests. Methods: The rotary nickel-titanium instruments used in this study were Reciproc and Mtwo. The instruments were initially subjected to a cantilever-bending test and then to static and dynamic fatigue tests. Reciproc instruments were operated in reciprocating movement, whereas Mtwo instruments were worked in continuous rotation. Results: The means of bending resistance (maximum load in grams) of the instruments were 274.9 for Reciproc and 429 for Mtwo. The mean times (in seconds) to fracture of the instruments subjected to static and dynamic tests were 214.5 (static) and 286.3 (dynamic) for Reciproc and 38.9 (static) and 99 (dynamic) for Mtwo. The Student's t test revealed significant differences in all tests (P < .05). Conclusions: The results of the present study showed longer fatigue life for instruments with higher flexibility, driven by reciprocating movement, and in the dynamic testing model. These findings reinforce the assumption that use of reciprocating movement is a means to prolong the fatigue life of rotary nickel-titanium endodontic instruments during instrumentation of curved canals.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available