4.7 Article

Phenotypic Detection of Carbapenemase-Producing Enterobacteriaceae by Use of Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry and the Carba NP Test

Journal

JOURNAL OF CLINICAL MICROBIOLOGY
Volume 52, Issue 11, Pages 4075-4077

Publisher

AMER SOC MICROBIOLOGY
DOI: 10.1128/JCM.02121-14

Keywords

-

Categories

Ask authors/readers for more resources

We compared the diagnostic accuracy of the Carba NP test with that of a straightforward matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) method for detecting carbapenemase-producing Enterobacteriaceae (CPE). Using PCR as the reference method, both tests demonstrated a sensitivity of 87% and a specificity of 100%. MALDI-TOF MS offers a potential alternative for the rapid detection of CPE in the clinical laboratory setting.