Determination of isothiazolinone preservatives in cosmetics and household products by matrix solid-phase dispersion followed by high-performance liquid chromatography-tandem mass spectrometry

Journal

JOURNAL OF CHROMATOGRAPHY A
Volume 1270, Issue -, Pages 41-50

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/j.chroma.2012.10.063

Keywords

Isothiazolinones; Cosmetics; Matrix solid-phase dispersion; Biocides; Preservatives; Personal care products; LC-MS/MS

Ask authors/readers for more resources

In this work, the development of a new efficient methodology applying, for the first time, matrix solid phase dispersion (MSPD) for the determination of sensitizer isothiazolinone biocides in cosmetics and household products - 2-methyl-3-isothiazolinone (MI), 5-chloro-2-methyl-3-isothiazolinone (CMI), 1,2-benzisothiazolinone (Bzl) and 2-octyl-3-isothiazolinone (OI) - is described. The main factors affecting the MSPD extraction procedure, the dispersive phase and the elution solvent, are assessed and optimized through a multicategorical experimental design, using a real cosmetic sample. The most suitable extraction conditions comprise the use of 2 g of florisil as dispersive phase and 5 mL of methanol as elution solvent. Subsequently, the extract is readily analyzed by HPLC-MS/MS without any further clean-up or concentration steps. Method performance was evaluated demonstrating to have a broad linear range (R-2 > 0.9980) and limits of detection (LOD) and quantification (LOQ) at the low nanogram per gram level, which are well below the required limits for UE regulation compliance. Satisfactory recoveries above 80%. except for MI (mean values close to 60%), were obtained. In all cases, the method precision (% RSD) was lower than 7%, making this low cost extraction method reliable for routine control. The validated methodology was finally applied to the analysis of a wide variety of cosmetics and household products. Most of the real samples analyzed have been shown to comply with the current European Cosmetic Regulation, although the results obtained for some rinse-off cosmetics (e.g. baby care products) revealed high isothiazolinone content. (c) 2012 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available