4.6 Article Proceedings Paper

Aspects of the Design Space in high pressure liquid chromatography method development

Journal

JOURNAL OF CHROMATOGRAPHY A
Volume 1217, Issue 19, Pages 3193-3200

Publisher

ELSEVIER
DOI: 10.1016/j.chroma.2010.02.001

Keywords

Robustness of HPLC methods; Method development; Optimization; Quality by Design; Design Space; 3D-Computer modelling software; Validation; QbD; ICH Q8

Ask authors/readers for more resources

The present paper describes a multifactorial optimization of 4 critical HPLC method parameters, i.e. gradient time (t(G)), temperature (T), pH and ternary composition (B-1:B-2) based on 36 experiments. The effect of these experimental variables on critical resolution and selectivity was carried out in such a way as to systematically vary all four factors simultaneously. The basic element is a gradient time-temperature (t(G)-T) plane, which is repeated at three different pH's of the eluent A and at three different ternary compositions of eluent B between methanol and acetonitrile. The so-defined volume enables the investigation of the critical resolution for a part of the Design Space of a given sample. Further improvement of the analysis time, with conservation of the previously optimized selectivity, was possible by reducing the gradient time and increasing the flow rate. Multidimensional robust regions were successfully defined and graphically depicted. (C) 2010 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available