4.5 Article

Low-cost three-dimensional imaging system combining fluorescence and ultrasound

Journal

JOURNAL OF BIOMEDICAL OPTICS
Volume 16, Issue 12, Pages -

Publisher

SPIE-SOC PHOTO-OPTICAL INSTRUMENTATION ENGINEERS
DOI: 10.1117/1.3662455

Keywords

fluorescence; US; reconstruction; imaging; tomography; prior information; artherosclerosis

Funding

  1. NSERC
  2. China Scholarship Council (CSC)

Ask authors/readers for more resources

In this paper, we present a dual-modality imaging system combining three-dimensional (3D) continuous-wave transillumination fluorescence tomography with 3D ultrasound (US) imaging. We validated the system with two phantoms, one containing fluorescent inclusions (Cy5.5) at different depths, and another varying-thickness semicylindrical phantom. Using raster scanning, the combined fluorescence/US system was used to collect the boundary fluorescent emission in the X-Y plane, as well as recovered the 3D surface and position of the inclusions from US signals. US images were segmented to provide soft priors for the fluorescence image reconstruction. Phantom results demonstrated that with priors derived from the US images, the fluorescent reconstruction quality was significantly improved. As further evaluation, we show pilot in vivo results using an Apo-E mouse to assess the feasibility and performance of this system in animal studies. Limitations and potential to be used in artherosclerosis studies are then discussed. (C) 2011 Society of Photo-Optical Instrumentation Engineers (SPIE). [DOI: 10.1117/1.3662455]

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available