Comparison of Acetabular Shell Position Using Patient Specific Instruments vs. Standard Surgical Instruments: A Randomized Clinical Trial

Journal

JOURNAL OF ARTHROPLASTY
Volume 29, Issue 5, Pages 1030-1037

Publisher

CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.arth.2013.10.006

Keywords

total hip arthroplasty; patient specific instruments; acetabular cup; preoperative planning

Categories

Ask authors/readers for more resources

Total hip arthroplasty (THA) survivorship relies largely upon appropriate acetabular cup placement. The purpose of this prospective randomized controlled trial was to determine whether the use of a preoperative 3D planning software in combination with patient specific instrumentation (PSI) results in improved cup placement compared with traditional techniques. Thirty-six THA patients were randomized into standard (STD) or PSI technique. Standard approach was completed using traditional techniques, while PSI cases were planned and customized surgical instruments were manufactured. Postoperative CT scans were used to compare planned to actual results. Differences found betweenplanned andactual anteversion were-0.2 degrees +/- 6.9 degrees (PSI) and -6.9 degrees +/- 8.9 degrees (STD) (P = 0.018). Use of 3D preoperative planning along with PSIs resulted in significantly greater anteversion accuracy than traditional planning and instrumentation. (C) 2014 Elsevier Inc. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available