4.7 Article

Improving acquisition times of elemental bio-imaging for quadrupole-based LA-ICP-MS

Journal

JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
Volume 27, Issue 1, Pages 159-164

Publisher

ROYAL SOC CHEMISTRY
DOI: 10.1039/c1ja10301f

Keywords

-

Ask authors/readers for more resources

Elemental bio-imaging experiments by quadrupole-based LA-ICP-MS normally employ scan speeds where the distance traversed in one second is equal to or less than the diameter of the laser beam. Consequently, data for a higher-resolution (pixel size = 15 mu m(2)) image of a 5 mm(2) tissue section can take upwards of 30 h to acquire. Appropriate laser scan speeds may be calculated by consideration of the relationship between laser scan speed, laser spot diameter and the total scan cycle of the quadrupole mass analyser. This paper presents a simple method to calculate the laser scan speeds capable of reducing the acquisition time by up to a factor of 5 whilst maintaining dimensional integrity of the image.