4.7 Article

Rapid Surface Plasmon Resonance Immunoassay for the Determination of Deoxynivalenol in Wheat, Wheat Products, and Maize-Based Baby Food

Journal

JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY
Volume 58, Issue 16, Pages 8936-8941

Publisher

AMER CHEMICAL SOC
DOI: 10.1021/jf101517s

Keywords

Deoxynivalenol; trichothecenes; cereals; screening assay; surface plasmon resonance

Funding

  1. European Commission [FOOD-CT-2005-006988 BIOCOP]

Ask authors/readers for more resources

A rapid screening assay (9 min/sample) has been developed and validated for the detection of deoxynivalenol in durum wheat, wheat products, and maize-based baby foods using an SPA biosensor. Through a single laboratory validation, the limits of detection (LOD) for wheat, wheat-based breakfast cereal, and maize-based baby food were 57, 9, and 6 mu g/kg, respectively. Intra-assay and interassay precisions were calculated for each matrix at the maximum and half-maximum European Union regulatory limits and expressed as the coefficient of variation (CV). All CVs fell below 10% with the exception of the between-run CV for breakfast cereal. Recoveries at the concentrations tested ranged from 92 to 115% for all matrices. Action limits of 161, 348, and 1378 mu g/kg were calculated for baby food, wheat-based breakfast cereal, and wheat, respectively, and the linear range of the assay was determined as 250-2000 mu g/kg.