Factors Predicting and Having an Impact on the Need for a Permanent Pacemaker After CoreValve Prosthesis Implantation Using the New Accutrak Delivery Catheter System

Journal

JACC-CARDIOVASCULAR INTERVENTIONS
Volume 5, Issue 5, Pages 533-539

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.jcin.2012.03.011

Keywords

aortic stenosis; pacemakers; prosthesis; transcatheter aortic valve implantation

Ask authors/readers for more resources

Objectives The purpose of this study was to evaluate the need for a permanent pacemaker after transcatheter aortic valve implantation with the CoreValve prosthesis (Medtronic, Inc., Minneapolis, Minnesota) using the new Accutrak delivery system (Medtronic, Inc.). Background The need for a permanent pacemaker is a recognized complication after transcatheter aortic valve implantation with the CoreValve prosthesis. Methods Between April 23, 2008 and May 31, 2011, 195 consecutive patients with symptomatic aortic valve stenosis underwent transcatheter aortic valve implantation using the self-expanding CoreValve prosthesis. In 124 patients, the traditional delivery system was used, and in 71 patients, the Accutrak delivery system was used. Results There were no significant differences in baseline electrocardiographic characteristics between the traditional system and the Accutrak patients: PR interval: 153 +/- 46 mm versus 165 +/- 30 mm, p = 0.12; left bundle branch block: 22 (20.2%) versus 8 (12.7%), p = 0.21; right bundle branch block: 21 (19.3%) versus 8 (12.7%), p = 0.26. The depth of the prosthesis in the left ventricular outflow tract was greater with the traditional system than with the Accutrak system (9.6 +/- 3.2 mm vs. 6.4 +/- 3 mm, p < 0.001) and the need for a permanent pacemaker was higher with traditional system than with Accutrak (35.1% vs. 14.3%, p = 0.003). The predictors of the need for a pacemaker were the depth of the prosthesis in the left ventricular outflow tract (hazard ratio [HR]: 1.2, 95% confidence interval [CI]: 1.08 to 1.34, p < 0.001), pre-existing right bundle branch block (HR: 3.5, 95% CI: 1.68 to 7.29, p = 0.001), and use of the traditional system (HR: 27, 95% CI: 2.81 to 257, p = 0.004). Conclusions The new Accutrak delivery system was associated with less deep prosthesis implantation in the left ventricular outflow tract, which could be related to the lower rate of permanent pacemaker requirement. (J Am Coll Cardiol Intv 2012;5:533-9) (C) 2012 by the American College of Cardiology Foundation

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available