Journal
INVESTIGATIVE RADIOLOGY
Volume 48, Issue 9, Pages 638-645Publisher
LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/RLI.0b013e31828a012d
Keywords
compressed sensing; clinical quality assessment; accelerated imaging; magnetic resonance neuroimaging
Ask authors/readers for more resources
Purpose: The aim of this study was to determine to what degree current compressed sensing methods are capable of accelerating clinical magnetic resonance neuroimaging sequences. Methods: Two 2-dimensional clinical sequences were chosen for this study because of their long scan times. A pilot study was used to establish the sampling scheme and regularization parameter needed in compressed sensing reconstruction. These findings were used in a subsequent blinded study in which images reconstructed using compressed sensing were evaluated by 2 board-certified neuroradiologists. Image quality was evaluated at up to 10 anatomical features. Results: The findings indicate that compressed sensing may provide 2-fold acceleration of certain clinical magnetic resonance neuroimaging sequences. A global ringing artifact and image blurring were identified as the 2 primary artifacts that would hinder the ability to confidently discern abnormality. Conclusion: Compressed sensing is able to moderately accelerate certain neuroimaging sequences without severe loss of clinically relevant information. For those sequences with coarser spatial resolution and/or at a higher acceleration factor, artifacts degrade the quality of the reconstructed image to a point where they are of little to no clinical value.