Comparison of kilovoltage cone-beam computed tomography with megavoltage projection pairs for paraspinal radiosurgery patient alignment and position verification

Journal

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.ijrobp.2008.04.029

Keywords

image-guided radiotherapy; cone-beam computed tomography; paraspinal; intrafractional motion; megavoltage localization

Funding

  1. NCI NIH HHS [P01 CA 59017] Funding Source: Medline

Ask authors/readers for more resources

Purpose: Implanted gold markers and megavoltage (MV) portal imaging are commonly used for setup verification of paraspinal tumors treated with high-dose, single-fraction radiotherapy. We investigated whether the use of kilo-voltage cone-beam computed tomography (CBCT) imaging eliminates the need for marker implantation. Methods and Materials: Patients with paraspinal disease who were eligible for single-fraction stereotactic body radiotherapy were accrued to an institutional review board-approved protocol. Each of 16 patients underwent implantation of fiducial markers near the target. The markers were visible on the MV images. Three MV image pairs were acquired for each patient (initial, verification, and final) and were registered to the reference images. Every MV pair was complemented by a CBCT scan. CBCT image registration was performed automatically by maximizing the mutual information using a region of interest that excluded the markers. The corrections, as determined from the MV images, were compared with these from CBCT and were used for actual patient setup. Results: The mean and standard deviation of the absolute values of the differences between the CBCT and MV corrections were 1.0 +/- 0.7, 1.0 +/- 0.6, and 1.0 +/- 0.8 min for the left-right, anteroposterior, and superoinferior directions, respectively. The absolute differences between the corresponding pre- and post-treatment kilovoltage CBCT image registration were 0.6 +/- 0.5, 0.6 +/- 0.5, and 1.0 +/- 0.8 mm. Conclusion: The setup corrections found using CBCT without the use of implanted markers were consistent with the marker registration on MV projections. CBCT has additional advantages, including better positioning precision and robust automatic three-dimensional registration, as well as eliminating the need for invasive marker implantation. We have adopted CBCT for the setup of all single-fraction paraspinal patients. Our data have also demonstrated that target displacements during treatment are insignificant. (C) 2008 Elsevier Inc.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available