FPGA Based Electronics for PET Detector Modules With Neural Network Position Estimators

Journal

IEEE TRANSACTIONS ON NUCLEAR SCIENCE
Volume 58, Issue 1, Pages 34-42

Publisher

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TNS.2010.2081685

Keywords

FPGA; neural network position estimators; positron emission tomography; signal readout and processing

Funding

  1. National Natural Science Foundation of China (NSFC) [10675120, 10975131]
  2. Specialized Research Fund for the Doctoral Program of Higher Education [20070358057]

Ask authors/readers for more resources

We are currently developing a prototype monolithic scintillator PET detector module based on neural network position estimators. The detector module comprises a 25.5 mm x 25.5 mm x 10 LYSO crystal coupled to a Hamamatsu 64 channels multi-anode PMT H7546B. The electronics for the detector module reads out all the signal channels, which represents the distribution of the scintillating light for gamma each event, and calculates the impinging position according to the pre-defined neural network algorithms if the event satisfies the energy and timing selection conditions. Compared with classical pixelated detectors, a monolithic scintillator based detector module features a simpler design, lower cost, and better energy resolution, but has lower signal to noise ratio and a more complicated signal readout scheme and data processing. By Monte-Carlo simulation, the performances of several readout schemes were compared. An optimized readout scheme which combines the 64 channels into 16 digitized signals was adopted in our electronics design. After the high resolution signal waveform digitization, an FPGA takes charge of the remaining digital signal processing, including the on-line hardware execution of the neural network positioning algorithms. We have implemented the electronics system for the detector modules. A pipelined implementation of the optimized neural network algorithms in the FPGA is able to process up to 15.3 M events per second without loss of performance compared to an off-line implementation. In addition to the function validation tests, the preliminary performance of the detector module we are building for a PET system is also reported.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available