Performance Evaluation of an MRI-Compatible Pre-Clinical PET System Using Long Optical Fibers

Journal

IEEE TRANSACTIONS ON NUCLEAR SCIENCE
Volume 57, Issue 3, Pages 1052-1062

Publisher

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TNS.2010.2044891

Keywords

Magnetic resonance imaging; optical fibers; PET; small animal imagers

Funding

  1. Engineering and Physical Sciences Research Council [GR/M22222/01]
  2. EPSRC
  3. PET Imaging Centre at St Thomas', St Thomas' Hospital
  4. Division of Imaging Sciences, King's College London

Ask authors/readers for more resources

We have designed and constructed an MR-compatible PET system for fully simultaneous PET/MR studies of small animals. The scanner uses long optical fibers to distance the magnetic field sensitive PET PMTs from the high magnetic field at the center of an MR scanner. It is a single slice system with an inner diameter of 7 cm. A full evaluation of the performance of the PET system and the results of an MR compatibility assessment in a Philips Achieva whole body 3 T MRI scanner are presented. The reconstructed resolution of the PET scanner is 1.5 mm at the center falling to 2.5 mm at the edge of the field of view; the system sensitivity is 0.95%; the count rate is linear up to an activity of 6 MBq (similar to 4 kcps) and the scatter fraction is 42% which can be reduced to 26% using MR-compatible gamma shields. Simultaneous PET/MR images of phantoms and a mouse have been acquired. The system is highly MR compatible, as demonstrated here, showing no degradation in performance of either the MR or PET system in the presence of the other modality. The system will be used to demonstrate novel pre-clinical applications of simultaneous PET/MR.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available