4.6 Article

Painless drug delivery through microneedle-based transdermal patches featuring active infusion

Journal

IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
Volume 55, Issue 3, Pages 1063-1071

Publisher

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TBME.2007.906492

Keywords

drug delivery; insulin; intradermal; microneedles; transdermal

Ask authors/readers for more resources

This paper presents the first microneedle-based transdermal patch with integrated active dispensing functionality. The electrically controlled system consists of a low-cost dosing and actuation unit capable of controlled release of liquid in the microliter range at low flow-rates and minimally invasive, side-opened, microneedles. The system was successfully tested in vivo by insulin administration to diabetic rats. Active infusion of insulin at 2 mu l/h was compared to passive, diffusion-driven, delivery. Continuous active infusion caused significantly higher insulin concentrations in blood plasma. After a 3-h delivery period, the insulin concentration was five times larger compared to passive delivery. Consistent with insulin concentrations, actively administered insulin resulted in a significant decrease of blood glucose levels. Additionally, insertion and liquid injection was verified on human skin. This study shows the feasibility of a patch-like system with on-board liquid storage and dispensing capability. The proposed device represents a first step towards painless and convenient administration of macromolecular drugs such as insulin or vaccines.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available