4.6 Article

Power-line interference detection and suppression in ECG signal processing

Journal

IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
Volume 55, Issue 1, Pages 354-357

Publisher

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TBME.2007.902234

Keywords

electrocardiogram (ECG); linear discriminant analysis (LDA); power-line interference (PLI); recursive least squares (RLS)

Ask authors/readers for more resources

A novel power-line interference (PLI) detection and suppression algorithm is presented to preprocess the electrocardiogram (ECG) signals. A distinct feature of this proposed algorithm is its ability to detect the presence of PLI in the ECG signal before applying the PLI suppression algorithm. No PLI suppression operation will be performed if PLI is not detected. We propose a PLI detector that employs an optimal linear discriminant analysis (LDA) algorithm to make a decision for the PLI presence. An efficient recursive least-squares (RLS) adaptive notch filter is also developed to serve the purpose of PLI suppression. Experimental results demonstrate superior performance of this proposed algorithm.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available